6
Error Information
Module Error Messages
Adapter Required
Error ID: 4961
The wellplate adapter is required for the following operation.
Probable cause
1
2
3
Funnel Not Supported
Error ID: 4962
Funnels are only supported for the analytical/bio-inert fraction collector.
Consequently the error message is displayed, if a tray with funnels is
configured for the preparative fraction collector.
Probable cause
1
106
Wellplate adapter not attached.
Cable between needle carrier assembly to
transport unit assembly disconnected or
defective.
Needle carrier assembly defective.
Fraction collector type and tray not
compatible.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Suggested actions
Attach wellplate adapter.
Check cable between needle carrier assembly
to transport unit assembly.
Exchange needle carrier assembly.
Suggested actions
• Check the tray configuration in your CDS.
Funnel trays are only supported for the
analytical scale fraction collector.
• Check the system configuration in your
CDS, and verify that an analytical scale
fraction collector is configured.