7
Maintenance
Exchange the Diverter Valve
Exchange the Diverter Valve
For bio-inert modules use bio-inert parts only!
When
If leaky or defective
Tools required
p/n
8710-2438
Parts required
p/n
G5664-60901
Preparations
• Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Turn OFF the instrument.
Connect the tubings as described in order to maximize their lifetime and operating
N O T E
security.
132
Description
Hex key 2.0 mm
Description
Diverter valve
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual