Using the Module

Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.

Materials in Flow Path (G7104C)

Following materials are used in the flow path of this module:
Part
Degasser chamber
Ultra clean tubings
Microfluidic
structures
MCGV
Passive inlet valve
Outlet valve
Pump head
Pistons
Piston/wash seals
Pressure sensor
Multipurpose Valve
1
Ultra clean tubings are available for the use with high-end MS detectors. They are also compatible to
THF.
2
Inlet Weaver, Jet Weaver, Heat Exchanger
1260 Infinity II Flexible Pump User Manual
Follow recommendations for avoiding the growth of algae, see
in HPLC Systems"
on page 70.
Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22 µm filters.
Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials like flow
cells, valve materials etc. and recommendations in subsequent sections.
Materials
TFE/PDD copolymer, PFA (internal tubings); PEEK (inlets); FEP (tubings);
ETFE (fittings)
PFA (tubings), PEEK (fittings)
1
SST
2
PEEK, FEP, PFA, Al
SST, gold, ruby, ZrO
SST, gold, ruby, ZrO
SST
ZrO
-based ceramic
2
UHMW-PE, SST
SST
Polyimide, SST, DLC
O
-based ceramic, ruby, sapphire, SST
2
3
-based ceramic, tantalum
2
-based ceramic, tantalum
2
"Algae Growth
58