Exchange the Diverter Valve
7
Install the valve to needle tubing at the new diverter
valve.
9
Connect the finger-tight fittings of the inner tubing kit to
the ports of the diverter valve. The tubings must run into
the ports of the diverter valve in horizontal, parallel lines.
N O T E
The tubings must not be bent or twisted.
134
8
Next Steps:
10
11
12
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
A color coded ring on one of the tubings and the valve
body indicates, which tubing belongs to which port.
Re-install the tray(s) in the tray base.
Start the instrument.
Close the front cover.