Replace The Valve To Needle Tubing (Analytical And Bio-Inert) - Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual

Infinitylab lc series
Table of Contents

Advertisement

7
Maintenance

Replace the Valve to Needle Tubing (Analytical and Bio-inert)

Replace the Valve to Needle Tubing (Analytical and Bio-inert)

For bio-inert modules use bio-inert parts only!
When
• When contaminated, worn out or visibly damaged
• Typically once every year
Tools required
p/n
8710-1534
8710-0510
Parts required
#
1
1
Preparations
• Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Position the transport unit of the fraction collector in the Change Parts position and turn off the
• It might be more convenient to remove the needle from its carrier before unscrewing the needle
Personal injury
WA R N I N G
Risk of personal injury caused by the needle arm movement.
122
Description
Wrench, open end, 4 mm
Wrench open 1/4 — 5/16 inch
p/n
G1364-68602
G1364-68601
instrument.
tubing.
Keep fingers away from the needle area.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Description
Tubing Kit 1 – 5 mL/min, 0.25 mm ID
Standard tubing kit for Analytical and Bio-inert Fraction Collector
Tubing Kit 1 mL/min, 0.15 mm ID (OPTIONAL)