Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual page 126

Infinitylab lc series
Table of Contents

Advertisement

7
Maintenance
Exchange the Analytical Needle Assembly
Liquid spills or fraction losses
C AU T I O N
Worn or damaged tubings can cause potential spills or lead to fraction losses.
Install the tubing as described to maximize its lifetime.
N O T E
1
Using the 4 mm wrench and the 5/16" wrench for
counter-holding, unscrew the valve to needle tubing
from the needle.
126
Explicitly follow the described installation procedures to maximize the lifetime of
inlet / waste tubing assembly and the valve to needle tubing, and to avoid potential
spills or fraction losses.
Regularly inspect the tubings and exchange them if they are worn out or show
visible signs of damage.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
2
Remove the wellplate adapter and the clip.