Preparing the Fraction Collector

Best Practices

Best Practices

Regular Inspections

Inspect the inlet/waste tubings and exchange them if they are worn out or
show visible signs of damage.

Power up / Shut down

Power up
• Check that the robotics is not obstructed.
Shut down
• Remove filled containers from the fraction collector after use.
• Pump a rinse solution through the fraction collector at the end of a run to
• Use recommended solvents to store the system.

Prepare the Fraction Collector

• Flush the LC system.
• Make sure to have a stable detector baseline.
• Make sure that fraction tubes are empty or that there is at least enough

Using the Fraction Collector

• Rinse the needle between runs.
• Pooling: Make sure that all fraction collection locations are large enough to
62
avoid clogging.
space for the next fraction.
completely collect all pooled fractions.