Using the Degasser

Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.

Material Information

Material in Flow Path
Following materials are used in the flow path of this module:
Table 3
Part
Internal tubings
Inlets
Tubings
Fittings
1260 Infinity II Degasser User Manual
Follow the recommendations for avoiding the growth of algae, see the pump
manuals.
Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.22 µm filters.
Avoid or minimize the use of solvents that may corrode parts in the flow path.
Consider specifications for the pH range given for different materials such as
flow cells, valve materials etc. and recommendations in subsequent sections.
Material in flow path
Material
PTFE
PEEK
FEP
ETFE