Maintenance
Exchange the Leak Sensor
5
Install the new leak sensor into its position in the leak
plane.
7
Re-install the front tray guide and fasten it with the three
screws. Be careful not to pinch the leak sensor cable.
140
6
Next Steps:
8
9
10
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Route the leak sensor cable nicely into the recess at the
front of the instrument. Plug the leak sensor into its
connector on the SLF board.
Re-install the tray(s) in the tray base.
Start the instrument.
Close the front cover.