Parts for Maintenance and Repair

List of Recommended Plates and Closing Mats

List of Recommended Plates and Closing Mats

Warnings
Explosive gas mixtures
WA R N I N G
There is a risk of explosive gas mixtures in the instrument if flammable solvents are
used.
Contamination with adhesives
C AU T I O N
Closing mats with adhesives can give some contamination in the system. The
adhesive is soluble in most of the solvents used in HPLC.
160
Cover the plates.
Remove the plates from the fraction collector after turning it 0FF.
Only use solvents with a flash point higher than 200 °C.
In general do not use closing mats with adhesive. The fraction collector has no
prepunch needle, therefore the adhesive will clog the needle after several
injections.