Error Information
Module Error Messages
Movement to Next Position Failed
Error ID: 4957
The transport mechanism detected an unexpected situation during the
movement to the next fraction position.
Probable cause
1
Could Not Find a Valid Next Position
Error ID: 4958
There has been more fractions than fraction positions.
Probable cause
1
2
110
Mismatch between tray configuration and
the loading of the trays, e.g. 4 well plates
are configured in the UI, but only three are
loaded, or shallow plates are configured,
but deep well-plates are used, or tube
height doesn't match the configuration.
Unexpected number of fractions.
Start of an analysis / sequence / run
without changing of the tray of the previous
run.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Suggested actions
Check the configuration and the loading of the
tray.
Suggested actions
• If possible use a tray with more positions.
• Add an additional fraction collector to the
system.
Ensure that there are enough fraction positions
for the complete analysis / sequence / run.