6
Error Information
General Error Messages

Leak Sensor Short

Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable cause
1
2

Leak Sensor Open

Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable cause
1
2
3
96
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Leak sensor not connected to the main
board.
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.