Preparing the Fraction Collector
Method Parameter Settings
Table 11
Method Parameter Settings
Parameter
80
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Description
Trigger Combinations
Use the Trigger Combinations to specify how
multiple peak triggers are combined to start or
stop Fraction Collection.
You can choose that:
• Collection of a fraction is started when all
peak detectors have sent a start trigger, and
continues until one detector sends a stop
trigger (AND condition)
• Collection of a fraction is started when at
least one peak detector has sent a start
trigger, and continues until all detectors send
a stop trigger (OR condition)
• Collection of a fraction is started when all
peak detectors have sent a start trigger, and
continues until all detectors send a stop
trigger (AND condition for start, OR condition
for stop)
Stoptime
Enables you to set a time at which the fraction
collector stops an analysis. If the fraction
collector is used with other Agilent Modular LC
modules, the fraction collector stoptime stops the
fraction collector only and does not stop any
other modules.
Limits: 0.01 – 99999.00 min or As Pump/Injector
Posttime
You can set the Posttime so that your fraction
collector remains in the post-run state during the
Posttime to delay the start of the next analysis.
When the Posttime has elapsed, the fraction
collector is ready for the next analysis.
Limits: 0.01 – 99999.00 min or Off (0.0 min)