Repair or Exchange a Funnel of the Internal Tray

When
When leaky or contaminated
Parts required
p/n
G1364-68730
G1364-43201
G1364-86708
5022-2200
Preparations
• Position the transport unit of the fraction collector in the Home position and remove all
• Turn OFF the instrument.
• Remove the internal tray, see
1
Turn the internal tray upside down, lift the flow delay
sensor (1.) and remove the funnel's waste tubing
through the flow delay sensor (2.).
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Description
Funnel seal kit (pack of 10)
Funnel coupler
Waste tubing kit 0.5T (analytical scale)
Funnel
installed trays from the tray base.

Repair or Exchange a Funnel of the Internal Tray

"Exchange the Internal Tray"
2
Remove the funnel assembly from the internal tray by
screwing it counter clockwise and lifting it out.
Maintenance
on page 135.
7
137