Compensation Sensor Open

Error ID: 0081
The ambient-compensation sensor (NTC) on the main board in the module has
failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If the
resistance across the sensor increases above the upper limit, the error
message is generated.
Probable cause
1

Compensation Sensor Short

Error ID: 0080
The ambient-compensation sensor (NTC) on the main board in the module has
failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If the
resistance across the sensor falls below the lower limit, the error message is
generated.
Probable cause
1
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Defective main board.
Defective main board.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.