Exchange the Analytical Needle Assembly

For bio-inert modules use bio-inert parts only!
When
• When the needle is visibly damaged
• When the needle is blocked or contaminated
Tools required
p/n
8710-1534
8710-0510
Parts required
#
1
1
Preparations
• Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Position the transport unit of the fraction collector in the Change Parts position and turn off the
• Remove the needle from its carrier before unscrewing the needle tubing.
Personal injury
WA R N I N G
Risk of personal injury caused by the needle arm movement.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Description
Wrench, open end, 4 mm
Wrench open 1/4 — 5/16 inch
p/n
G1367-87200
G5667-87200
instrument.
Keep fingers away from the needle area.

Exchange the Analytical Needle Assembly

Description
Analytical needle assembly