Exchange the Needle/Capillary Carrier Assembly (Analytical and Bio-inert)

Exchange the Needle/Capillary Carrier Assembly (Analytical
and Bio-inert)
When
If defective
Tools required
p/n
8710-2438
Parts required
#
1
Preparations
• Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Position the transport unit of the fraction collector in the Change Parts position and turn OFF
• Remove the needle from its carrier before unscrewing the needle tubing.
Personal injury
WA R N I N G
Risk of personal injury caused by the needle arm movement.
Liquid spills or fraction losses
C AU T I O N
Worn or damaged tubings can cause potential spills or lead to fraction losses.
128
Description
Hex key 2.0 mm
p/n
G1364-60022
the instrument.
Keep fingers away from the needle area.
Explicitly follow the described installation procedures to maximize the lifetime of
inlet / waste tubing assembly and the valve to needle tubing, and to avoid potential
spills or fraction losses.
Regularly inspect the tubings and exchange them if they are worn out or show
visible signs of damage.