Maintenance
Replace the Inlet/Waste Tubings (Analytical and Bio-inert)
11
Install the leak adapter to the module housing.
13
Clip the inlet / waste tubing assembly into the leak
adapter.
15
Connect the finger-tight fittings of the inner tubing kit to
the ports of the diverter valve. The tubings must run into
the ports of the diverter valve in horizontal, parallel lines.
N O T E
The tubings must not be bent or twisted.
120
12
14
16
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Connect the outer tubing kit to the unions.
A color coded ring on one of the tubings and the valve
body indicates, which tubing belongs to which port.
Connect the tubing with the label with the arrow (1) to
the flow cell of the detector. Insert the other tubing (2)
into the waste container.