5
Remove the leak adapter from the module housing.
7
Remove the clip by pushing it down through the module
housing.
9
Slide in the new inner tubing kit from bottom to top (1.)
and let the clip click into position (2.).
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Replace the Inlet/Waste Tubings (Analytical and Bio-inert)
6
Remove the unions from the inner tubing kit.
8
Remove the inner tubing kit by pulling it down through
the opening in the module housing.
10
Install the unions to the inner tubing kit.
7
Maintenance
119