Pusher Wrong or Defect
Error ID: 4965
Probable cause
1
2
Wrong or Missing Needle (Analytical Scale)
Error ID: 4966
Probable cause
1
2
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Wellplate adapter not attached.
Needle carrier assembly defective.
No needle installed.
Short needle for high flow rates installed,
but the wellplate adapter hasn't been
removed.
Error Information
Module Error Messages
Suggested actions
Attach wellplate adapter.
Exchange needle carrier assembly.
Suggested actions
Check which needle has been installed.
Remove wellplate adapter if short needle for
semi-preparative operation is installed.