Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual page 129

Infinitylab lc series
Table of Contents

Advertisement

1
Using the 4 mm wrench and the 5/16" wrench for
counter-holding, unscrew the valve to needle tubing
from the needle.
3
Un-clip the tubing from the bottom of the needle carrier
assembly.
5
Unplug the blue connector of the brown flat band cable
of the needle carrier assembly from its plug at the z-arm.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Exchange the Needle/Capillary Carrier Assembly (Analytical and Bio-inert)
2
Hold the needle/capillary guiding assembly between
your thumb and forefinger and slide out the assembly
towards the rear of the needle carrier assembly.
4
Slide the tubing through the hole in the needle carrier
assembly (from bottom to top) (1.) and out of the holder
in the z-arm assembly (2.).
6
Locate the three holding screws of the needle carrier
assembly and unscrew them.
7
Maintenance
129