4
Preparing the Fraction Collector
Example of a perfect fitting
5
Once you believe you have the fitting complete, loosen
the nut, and inspect the ferrule for the correct position
on the tubing.
Example of a perfect fitting
Examples of incorrect fittings
Figure 11
Examples of incorrect fittings
The first time that the swagelock fitting is used on a column or an injection valve, the
N O T E
position of the ferrule is permanently set. If changing from a column or an injection valve
to another, the fitting may leak or decrease the quality of the separation by contributing to
band broadening.
72
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual