Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual page 18

Infinitylab lc series
Table of Contents

Advertisement

1
Introduction
Bio-inert Materials
Table 1
Used Bio-inert materials
Module
Bio-inert Heat Exchangers, Valves and Capillaries:
Quick-Connect Heat Exchanger Bio-inert (G7116-60009)
(for Agilent 1260 Infinity II Multicolumn Thermostat G7116A)
Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A)
Bio-inert Connection capillaries
To ensure optimum bio-compatibility of your Agilent 1260 Infinity II Bio-inert LC system, do
N O T E
not include non-inert standard modules or parts to the flow path. Do not use any parts that
are not labeled as Agilent "Bio-inert". For solvent compatibility of these materials, see
"Material Information"
18
on page 64.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Materials
PEEK (steel-cladded)
PEEK, ceramic (Al
O
based)
2
3
Upstream of sample introduction:
• Titanium
Downstream of sample introduction:
• Agilent uses stainless-steel-cladded
PEEK capillaries, which keep the flow
path free of steel and provide
pressure stability to more than
600 bar.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1260 infinity ii analyticalfraction collector

Table of Contents