Physical Specifications

Table 2
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating
temperature
Ambient non-operating
temperature
Humidity
Operating altitude
Non-operating altitude
Safety standards:
IEC, EN, CSA, UL
ISM Classification
Permitted solvents
1
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Physical Specifications
If a thermostat is used the upper value for humidity can be reduced. Please check your lab condi-
tions to stay beyond dew point values for non–condensing operation.
Site Requirements and Specifications
Specification
13.5 kg (29.8 lbs)
200 x 345 × 440 mm
(8 x 13.5 × 17 inches)
100 – 240 V~, ± 10 %
50 or 60 Hz, ± 5 %
200 VA / 180 W
4 – 40 °C (41 – 104 °F)
-40 – 70 °C (-40 – 158 °F)
< 95 %, at 25 – 40 °C (77 – 104 °F)
Up to 3000 m (9842 ft)
Up to 4600 m (15092 ft)
Installation category II, Pollution degree 2
ISM Group 1 Class B
Boiling point ≥56 °C
Auto-ignition temperature ≥200 °C
Ignition Class IIA, IIB (IEC60079-20-1)
Physical Specifications
Comments
w/o Thermostat
Wide-ranging
capability
1
Non-condensing
For storing the module
For indoor use only.
According to CISPR 11
2
23