Needle to Needle Rinse / Funnel Position Failed

Error ID: 4955, 4980, 4981-4990
The needle failed to reach the needle rinse / funnel position.
The position of the needle is monitored by a position encoder on the needle
carrier. If the needle fails to reach the end point, or if the encoder fails to
recognize the needle carrier movement, the error message is generated.
Probable cause
1
2
3
4
5
6
7
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Bad sample transport unit alignment
Bent needle.
Missing needle.
Blocked rinse or funnel position.
Defective needle carrier assembly.
Disconnected needle carrier connector.
Defective MTP main board.
Error Information
Module Error Messages
Suggested actions
Do a self-alignment
Check and exchange the needle assembly if
necessary.
Check and exchange the needle assembly if
necessary.
Clean or change the funnel or rinse port
assembly if necessary.
Exchange the needle carrier assembly.
Connect needle carrier connector correctly.
Please contact your Agilent service
representative.