Table 12
Advanced Parameters Description
Parameter
3rd Party Pump Flow
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Preparing the Fraction Collector
Method Parameter Settings
Description
Delay Settings
Use the Delay Settings table to specify the delay
that is applied to a peak trigger signal. You can
specify this setting for each peak detector
separately. You can choose from:
• Off (No delay is applied to fraction collection
and collection starts as soon as the trigger
conditions are met)
• As calibrated (Delays fraction collection by a
pre-defined delay volume, where for each
peak trigger, the delay volume can be
displayed (and edited) using the Modify
Detector Delay Volumes dialog box, accessed
from the context menu of the instrument's
dashboard panel)
• Use Time (Enables the Time field to allow you
to set a delay time)
• Use Volume (Enables the Volume field to
allow you to set a delay volume)
Delay end of fraction: An additional delay can be
set if you want to delay the end of fraction
collection by an additional amount of time.
Specify the additional time used to delay the end
of fraction collection in seconds.
Fill Volume Settings
Use the Fill Volume Settings to specify the
Maximum fill volume used in your method.
If your Fraction Collector is not connected to a
Linked Pump, specify a Pump Flow for the
Fraction Collection method.
N O T E
This section is only visible if there is no
Agilent pump recognized.