Error Information
Module Error Messages

Initialization Failed

Error ID: 4950
The fraction collector failed to complete initialization correctly. The fraction
collector initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined routine. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable cause
1
2
3
4
5
108
Transport unit not aligned correctly
Mechanical obstruction.
Defective transport assembly flex board.
Defective MTP main board.
Excessive weight on top of the fraction
collector (see also
"Side Door Error"
page 101 )
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Suggested actions
Perform an auto-alignment.
Ensure unobstructed movement of the
transport assembly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Check stack configuration and reduce weight
on top of the fraction collector.