Perfom a Delay Calibration Run in OpenLAB CDS Chemstation
Edition
The delay calibration procedure determines the delay time between
detector(s) and the fraction collector in the system. The delay is used to
compensate for the time a compound needs to travel between the point of
detection in the detector and the point of collection in the fraction collector.
The delay calibration procedure is performed using the flow delay sensor
(FDS), a very simple detector built into the fraction collector. Together with
the signal from the detector, the signal from the FDS facilitates determination
of the delay between detector and fraction collector.
The figure shows the scheme of the delay time calculation between UV
detector and fraction collector.