Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual page 133

Infinitylab lc series
Table of Contents

Advertisement

1
Locate the diverter valve with the finger tight fittings of
the inner tubing kit.
3
Unscrew the finger-tight fitting of the valve to needle
tubing assembly at the diverter valve.
5
Using the 2 mm hex key loosen the holding screw of the
diverter valve and remove the diverter valve from the
z-arm.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Exchange the Diverter Valve
2
Unscrew the two finger-tight fittings of the inner tubing
kit at the diverter valve.
4
Disconnect the power cable of the diverter valve from
the connector on the front of the z-arm.
6
Install the new diverter valve to the z-arm with its
holding screw. Plug in the new diverter valve power
cable to the z-arm.
7
Maintenance
133