Vessel Stuck to Needle
Error ID: 4453
The vessel sticks to the needle when the needle moves up.
Probable cause
1
2
3
Cluster Partner Lost During Analysis
Error ID:
There was a problem with the inter module communication.
Probable cause
1
2
3
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Closing mat to rigid/thick.
Bad X or Theta positioning and the needle
sticks into the wall between two holes.
Defective encoder on the needle carrier
assembly.
Disconnected or defective CAN cable.
Disconnected or defective
24V-CAN-DC-OUT cable for an external
valve.
Defective UIB, external Valve or MTP board.
Error Information
Module Error Messages
Suggested actions
Check that the closing mat is not too thick.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Check the interconnection between the
modules.
Reconnect the UIB / Valve. Start a test
analysis/run.
• Switch power off / on (complete system
off, then on). Start a test analysis / run.
• Please contact your Agilent service
representative.