Perform Calibration Run
Use the Perform Calibration page of the Delay Calibration Wizard to start the delay
calibration run using the current method. The run parameters are listed; a
warning is given if the method has been changed. You can switch to a different
method or edit the sample information before starting the run.
Edit Sample Info
Load Method
Automatically activate Delay
Sensor Signal
Start Calibration Run
Delay Evaluation
At the end of each calibration run, you can choose to either evaluate the data
or start another calibration run.
Back
Cancel
Next
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
System On
System Off
Configuration and Operation of the Fraction Collector
Click to turn the system on.
Click to turn the system off.
Displays the Sample Info dialog box, which allows you to edit the
sample information for the calibration run.
Displays the Method Browser for master methods, which allows
you to select a different master method to load and use for the
calibration run.
Mark this check box to automatically switch on the collection of
the signal from the fraction collector's flow delay sensor. This
signal is necessary to calculate the delay time/volume.
N O T E
When you mark this check box, the method is modified.
Starts the delay calibration run. The message line describes the
progress of the run.
You can perform multiple calibration runs; this button is active
after each calibration run has completed.
Opens the Delay Evaluation window to allow you to determine
the delay volume(s)
("Delay Evaluation"
This button is active only when at least one delay calibration run
has been performed.
Displays the Prepare Instrument for Calibration page of the Delay
Calibration Wizard
("Prepare Instrument for Calibration"
page 36).
This button is active only until a calibration run has been started.
Displays the
"Delay Calibration - Close"
Displays the Finalize Calibration page of the Delay Calibration
Wizard
("Finalize Calibration"
This button is inactive until the calibration run is complete.
Using the Module
on page 40).
on page 38 dialog box.
on page 38).
3
on
37