Table 3
Performance Specifications 1260 Infinity II Analytical Fraction Collector (G1364F)
Type
Specification
Operating Modes
Discrete fractions: default mode for all vessels.
The flow is diverted to waste, while moving from one vessel position to the
next vessel position
Continuous flow: optional, available only when using deep well plates.
It is possible to move from one well plate position to the next one without
diverting the flow into the well plate to waste
Needle into location:
Needle pushes into the vessel as deep as specified, for the use with
capped vials and test tubes and well plates with closing mats
Droplet setup mode:
enables the fraction collector to collect small fractions without bubbles.
The tip of the fraction collector needle initially moves down to the bottom
of the well. Then it slowly moves upwards while the fraction is collected.
Maximum time to
Movement in x-direction: < 0.15 s
move between
Movement in y-direction: < 0.3 s
neighboring vessels
Diverter valve
3/2 valve, with switching time < 100 ms
Maximum pressure
6 bar at the diverter valve during switching
Plates/Trays
4 x well-plates full tray (MTP)
2 × well-plates std. tray + 10 funnels with external containers
tray)
2 × well-plates std. tray (MTP) + 10 × 2 mL vials
100 x 2 mL in std. tray (+ 1 half tray)
3 x 40 x 2 mL in half tray
3 x 40 funnels in half tray
3 x 15 x 6 mL in half tray
Full tray with 40 test tubes (30 mm OD, max. height 48 mm, ~20 mL / tube)
Full tray with 60 test tubes (25 mm OD, max. height 48 mm, ~15 mL / tube)
Full tray with 126 test tubes (16 mm OD, max. height 48 mm, ~11 mL / tube)
Full tray with 215 test tubes (12 mm OD, max. height 48 mm, ~8 mL / tube)
Installed trays are automatically detected and identified.
In operation mode "Needle into location" installed plates and vials can be
detected.
Only one type of well-plates can be used at a time in one tray.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
1
1
(+ 1 half tray)
1
1
1
Site Requirements and Specifications
Performance Specifications
1
(+ 1 half
2
Comment
25