Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual page 39

Infinitylab lc series
Table of Contents

Advertisement

Recommended Calibrants
Table 5
Fraction Collector type
G1364 B/C/D/E/F, G7159B and
G7166A
G5664 A/B
Table 6
Calibrants and recommended wavelength settings for UV-based delay calibration
Part No.
Name
5190-8223
Agilent delay and checkout calibrant
5190-6887
Prep LC Standard #2
G1946-85020
Delay Sensor Calibration Sample Kit
Table 7
Calibrants and target masses for MSD delay calibration
Part No.
Name
5190-8223
Agilent delay and checkout
calibrant
5190-6887
Prep LC Standard #2
G1946-85020
Delay Sensor Calibration Sample
Kit
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Configuration and Operation of the Fraction Collector
Recommended calibrants (P/N) for Agilent Fraction Collectors using
ChemStation
Dye
Patent Blue VF
Thionin acetate
FD&C Green 3 (Caffeine for MSD)
Dye
Patent Blue VF
Thionin acetate
FD&C Green 3 (for UV and
fraction delay sensor)
Caffeine (for MSD)
UV
UV and MSD
5190-8223
5190-8223
G1946-85020 or
5190-6887 or
5190-8223
5190-8223
Using the Module
Wavelength
600 nm
600 nm
600 nm
Target Mass
m/z 543.1 (neg)
m/z 545.1 (pos)
m/z 589.1 (pos, Na adduct)
m/z 228.1 (without acetate)
m/z 195.1 (pos)
3
39