Setting The 6-Bit Configuration Switch - Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual

Infinitylab lc series
Table of Contents

Advertisement

10
Hardware Information

Setting the 6-bit Configuration Switch

Setting the 6-bit Configuration Switch

The 6-bit configuration switch is located at the rear of the module with
FUSION electronics. Switch settings provide configuration parameters for
LAN and instrument specific initialization procedures.
All modules with FUSION electronics:
• Default is ALL switches DOWN (best settings).
• For specific LAN modes switches 4-5 must be set as required.
• For boot resident/cold start modes switches 1+2 or 6 must be UP.
Figure 19
192
Default IP address for LAN 192.168.254.11
Location of Configuration switch (example shows a G7114A/B VWD)
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual