Performance Specifications (G5664B)
Table 4
Performance Specifications 1260 Infinity II Bio-inert Fraction Collector (G5664B)
Type
Specification
Delay Volume (in µL)
Fraction collector inlet to diverter valve: ~50 (typical, depends on length of
the tubing)
Diverter valve: ~15
Diverter valve to needle: ~10
Needle: ~4
Minimum system flow Depending on the recommended flowrates of the installed tubing kit
Maximum system
10 mL/min
flow
Maximum collection
~20 mL with 30 x 48 mm (OD x L) tube
volume
Maximum capacity
3 fraction collectors in parallel plus one recovery fraction collector
Cooling
Optional (with additional G1330B), performance depending on ambient
conditions and the volume of collected fractions
Trigger modes
Time slices
Peak (threshold, up- / downslope)
Timetable (combination of time intervals and peak)
Manual trigger (supported only with Agilent Instant Pilot G4208A)
Trigger Sources
G7115A, 1260 Infinity II DAD
G7165A, 1260 Infinity II MWD
G7114A, 1260 Infinity II VWD
G6120BA, LC/MS Single Quad VL
G6130BA, LC/MS Single Quad SL
G7121A, 1260 Infinity II FLD
G4260B, 1260 Infinity II ELSD
G7162A, 1260 Infinity II RID
Other detectors can be used but are not supported for fraction collection.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Site Requirements and Specifications
Performance Specifications
2
Comment
LC and CE drivers
A.02.19 (or above) are
required for clustering
27