To enter customer service mode

A
B
C
D
1 Turn controls (C) and (D) VT and RR all the way
to the right.
2 Switch ON (E) the device (briefly press
and simultaneously press and hold the Values
key (A) and the Values
the main Customer Service Mode menu
appears.
3 Set the number of the required test in the main
menu with the central rotary knob.
4 Activate test = press rotary knob.
Instructions for Use Oxylog 2000 plus SW 1.n
Settings in customer service mode
1 Select the required function with the cursor
(asterisk).
– To select the parameter: turn the rotary
– To activate the parameter: press the rotary
Oxylog 2000 plus
– To set the value: turn the rotary knob.
– To confirm the value: press the rotary knob.
E
To exit the parameter settings menu
O
key)
key (B) until
1 Select the line EXIT.
2 Press the rotary knob and confirm.
The set values are saved and remain effective
whenever ventilation is started after switching
ON.
knob.