NIV – Non-invasive ventilation
Mask ventilation (optional)
NIV can only be activated as a supplementary
function in the pressure-controlled ventilation
modes SpnCPAP and SpnCPAP / PS. Mask leak-
ages are detected by the device, compensated and
included in the measured values for VTe and MVe.
WARNING
If NIV is not activated, measured values for VTe
and MVe will be inconsistent if there are
leakages during ventilation.
Use of NIV
WARNING
Dead space increases when using masks.
Note the mask manufacturer's instructions!
WARNING
Application mode NIV may not be activated
with intubated patients!
Risk of undetected leaks and inadequate ven-
tilation!
WARNING
Check MV alarm limits after deactivating NIV
mode!
WARNING
Avoid high airway pressure.
Risk of aspiration!
To switch on NIV
1 Press the Settings
appears.
2/3
2 Activate the line NIV off.
3 Select NIV on and confirm.
54
key until display page
– The supplement NIV appears in the upper
section of the display.
Oxylog 2000 plus automatically adjusts to the
requirements of mask ventilation. Leakage flows
are compensated automatically and the leakage
alarm is inactive.
WARNING
Set the lower alarm limit MV
the minimum ventilation required for the
patient.
Otherwise, there is a risk of the patient
receiving insufficient ventilation.
Apnea ventilation is not permitted by the ventilator
when NIV is active.
Instructions for Use Oxylog 2000 plus SW 1.n