Connecting the power supply

The Oxylog 2000 plus is designed to operate on
power supplies with different voltages.

Internal supply

– With rechargeable battery (specified Smart
Battery, refer to the "Technical Data"
on page 103 for additional information).

Additional external power supply

To recharge the battery and to extend the electrical
operation time.
– DC voltage from the on-board power supply via
DC/DC converter or

Internal supply with rechargeable battery

Replacing the battery

A
B
C
32
– with AC/DC power pack
WARNING
A fully charged battery must always be
installed for safety reasons, even when oper-
ating from an external power supply!
To have a fully charged battery on hand, refer to the
"Getting started" section on page 38 for additional
information.
WARNING
Treatment of batteries:
– Do not throw into fire,
– Do not force open,
Danger of bodily injury.
1 Loosen the screw (C) on the battery compart-
ment cover (B) counterclockwise to release the
cover.
2 Remove the battery cover.
3 Remove the battery (A) by pulling the tab.

Checking the charge of the battery

Press the button on the rechargeable battery.
The charge status is indicated as a percentage
by LEDs.