Technical Data - Dräger Evita 2 Instructions For Use Manual

Intensive care ventilator
Table of Contents

Advertisement

Technical Data
I
Environmental
conditions
For operation:
temperature
atmospheric pressure
rel. humidity
For storage:
temperaiure
atmospheric pressure
rel. humidity
Set values
Ventilation frequency for IPPV
Ventilation frequency for IMV
Ventilation frequency for BIPAP
I:E ratio TI:TE
Tidal volume VT
paediatric
accuracy
adult
accuracy
lnspiratory flow Vmti
paed.
adult
Inspiratory pressure;limitation
pm=
02 concentration
accuracy
Positive endexspiratory
pressure
PEEP or continuous positive airway
pressure CPAP or interm. PEEP
Trigger sensitivity A ptr
(differential pressure below PEEP)
for IPPV
and
for Slh4V
\
\
i
15 to 40°C
530 to 1100 hPa
0 to 90%
- 20 to 60°C
530to
1lOOhPa
0 to 90%
5 to 1OOImin
0 to 60lmin
2 to 1001min
6 : 1 to 1 : 6
0.04 to 0.3 L
2 10 % of set value or
f 70 mL, whichever the greater
value.
0.1 to 2.0 L
k 10% of set value or
t 25 mL, whichever the greater
value.
6 to 30 Umin
6 to 120 Llmin
0 to 1OO mbar
21 to lrxl vol.%
* 5% of set value or f 2 vol.%
whichever the greater value
0 to 35 mbar
0.7 mbar, fixed trigger

Advertisement