M300/M300+ Performance Data - Dräger Infinity M300 Series Instructions For Use Manual

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

Technical specifications
Electrical specifications
Input Voltage
Input frequency (Hz)
Protection Class
Mode of Operation
Swappable battery recharging time
Environmental requirements
Atmospheric pressure
Temperature
Humidity (non-condensing)
Protection against ingress of water

M300/M300+ Performance data

The following disclosures related to algorithm performance using certain waveforms are added or updated
to the monitoring specifications:
M300/M300+ ECG/Arrhythmia/ST Supplemental information
Time to alarm for tachycardia
Tall T wave rejection capability
Heart rate averaging method
284
Infinity M300+ battery charging station specifications
100 V to 240 V AC
CAUTION:
Use a surge protector with all electrical hookups.
50 Hz to 60 Hz AC inputs
Class 1
Continuous
0 to 100 % within approximately 4 h
Operating: 65 to 106 kPa
Storage: 50 to 106 kPa
Operating: 0 to 40 °C (32 to 104 °F)
Storage: –20 to +60 °C (–4 to +140 °F)
Operating: 10 to 95 %
Storage: 10 to 95 %
IPX1 (dripping) per IEC 60529
Ventricular tachycardia 1 mVpp, 206 bpm
Gain 0.5, range 3.0 to 3.5 seconds, average 3.3 seconds
Gain 1.0, range 2.9 to 3.3 seconds, average 3.2 seconds
Gain 2.0, range 2.8 to 3.5 seconds, average 3.0 seconds
Ventricular tachycardia 2 mVpp, 195 bpm
Gain 0.5, range 2.2 to 4.0 seconds, average 3.0 seconds
Gain 1.0, range 1.9 to 2.5 seconds, average 2.3 seconds
Gain 2.0, range 2.0 to 2.9 seconds, average 2.5 seconds
1.20 mV T wave amplitude
Heart rate is normally based on the average R-R interval calculated
over the last 10 seconds.
Instructions for use – Infinity
®
M300 and M300+ series – VG3.0