Ambient Conditions; Machine Data - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Technical Data

Ambient Conditions

During operation
Temperature
Atmospheric pressure
Relative humidity
During storage
Temperature
Atmospheric pressure
Relative humidity
The service conditions for supplementary equipment must be noted. These may restrict the area of use
for the system as a whole. Vaporizer units and the anesthetic agents used may restrict the area of use of
the workstation in regards to its temperature range and maximum fresh-gas flow. The corresponding
Instructions for Use of the supplementary equipment must be noted.

Machine Data

Gas supply from medical gas pipeline system
Pipeline System Pressure Range at Machine Connector
O
, N
O, AIR:
2
2
Gas supply connectors:
Each inlet is fitted with a non-return valve
Pipeline Pressure Indicator Accuracy
Gas supply from supplementary O
Cylinder Connections
Cylinder Gas Pressure
(typical full loads at 70 ºF, 21 ºC)
Cylinder Gauges
Cylinder Gauge Range
176
50 to 95 °F (10 to 35 °C)
700 to 1060 cmH
20 to 80 % (no condensation)
14 to 140 °F (–10 to 60 °C)
700 to 1060 cmH
10 to 90 % (no condensation)
50 to 55 psi (3.4 to 3.8 kPa x 100)
Note: Pipeline system supply pressure variation shall
not exceed ±10 %
DISS
±3 % of full scale from 40 to 120 psi
(2.7 to 8.3 kPa x 100)
AIR and N
O cylinders (with pin-index connections)
2,
2
Pin-indexed hanger yokes (CGA V-1-1994)
O
, AIR
2
N
O
2
Conform to ASME B40.1 Grade B
O
2
N
O
2
AIR
O (hPa)
2
O (hPa)
2
1900 psi (131 kPa x 100)
745 psi (51.3 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
0 to 3000 psi (206.8 kPa x 100)
Instructions for Use Fabius OS SW 3.n