Messages in the information window

Message
RR = 12 per min or
VT = 800 mL
I : E = 1 : 1.5
Flow = 15 L/min
RR = 12 per min or
VT = 800 mL
Ti = 0.7 s
Flow = 35 L/min
I : E = 1 : 1.5
Te = 2 s
Confirm PEEP
above 10 mbar ?
Gas
consumption = 10 L/min
(Battery capacity)
Psupp = 22mbar
Cause
Change in Ti, RR or VT
in ventilation mode VC-SIMV.
Change in I:E, RR or VT in
ventilation mode VC-CMV, VC-AC.
Change in Ti.
PEEP >10 mbar has been set but not confirmed.
Standard display in information window
for the current gas consumption.
Standard display in information window for the
current battery capacity.
Δ
Change in
Psupp or PEEP.
Explanation/Remedy
The required setting of PEEP >10 mbar is only
possible when confirmed via the rotary knob.
Psupp is the absolute pressure resulting from
Δ
PEEP +
Psupp.