Millipore Spectroquant Multy Operating Manual page 86

Colorimeter
Table of Contents

Advertisement

Measuring range: 0.02 – 4.50 mg/l Cl
Check the pH of the
Select method 1 3 3 ,
sample, specified range:
select subitem >>free.
pH 4 – 8.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Add 10 ml of the sample
Reaction time:
with pipette, close with
1 minute
the screw cap, and mix.
Press Enter to start
the countdown.
Important:
Very high chlorine concentrations in the sample produce
yellow-colored solutions (measurement solution should be
red) and false-low readings are yielded. In such cases the
sample must be diluted (plausibility check).
Release 08/2014 - Spectroquant
Chlorine
(with liquid reagents)
Detemination of free chlorine
24-mm cell
2
Fill approx. 10 ml of
distilled water into a
24-mm cell (do not add
any reagents!), close
with the screw cap.
(Blank cell)
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .
®
Multy
Place 6 drops of Cl
into a 24-mm cell.
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) a freshly prepared standard
solution can be used (see section "Standard solutions").
100086/100087
Test
-1
Add 3 drops of Cl
-2,
2
2
close with the screw
cap, and mix.