Millipore Spectroquant Multy Operating Manual page 195

Colorimeter
Table of Contents

Advertisement

Measuring range: 1 – 100 FAU
Select method 5 2 0 .
Fill approx. 10 ml of
distilled water into a
24-mm cell, close with
the screw cap.
(Blank cell)
Release 08/2014 - Spectroquant
Turbidity
analogous to EN ISO 7027
24-mm cell
Transfer the sample into
a 24-mm cell, close with
the screw cap.
®
Multy
Insert the blank cell into
Insert the cell containing
the cell compartment.
the sample into the cell
Align the mark on the
compartment. Align the
cell with that on the pho-
mark on the cell with
tometer. Press Zero .
that on the photometer.
Press Test .