Millipore Spectroquant Multy Operating Manual page 262

Colorimeter
Table of Contents

Advertisement

The data for the other coefficients are prompted in
sequence (B, C, D, E and F).
Note
Entering zero [0] for the value of a given coefficient
automatically negates any entry of the exponent.
Confirm each selection by pressing [Enter].
The display now shows:
Enter the measurement-range limits in the range between -2600
and +2600 mAbs.
• Press the arrow key [] or [] to select between the plus
and minus signs.
• Enter the lower limit (Min) and the upper limit (Max) in the
unit absorbance (mAbs),
e.g.: [2] [1] [0] [0] for 2100 mAbs.
Confirm each selection by pressing [Enter].
The display now shows:
Press the arrow key [] or [] to select the desired unit.
Confirm your selection by pressing [Enter].
measurement range
Min mAbs: +____
Max mAbs: +____
choose unit:
>> mg/l
g/l
mmol/l
mAbs
μg/l
E
A
%
Chapter 5.6 | Page 9 of 13