Millipore Spectroquant Multy Operating Manual page 267

Colorimeter
Table of Contents

Advertisement

5.7
User-specific calibration
In principle it is possible for the user to make his/her own
calibration. It is, however, advisable to retain the factory
calibration, since this was performed using a 10-item
cali bration procedure.
A user-specific calibration is made using a standard with a
known concentration. This concentration should be equivalent
to that of the water sample. Here it is possible to use e.g.
Spectroquant® CombiCheck standards or ready-to-use standard
solutions (see chapter 5.2).
In the case of differentiated methods, only the simple form is
calibrated, i.e. with the chlorine methods only free chlorine
is calibrated, and the calibration then automatically applies for
the other two variants (total and differentiated).
The following methods cannot be user-specifically calibrated:
Method No.:
10
20
21
70
90
110
140
170
560
240
270
290
300
322
550
555
350
360
400
401
410
440
441
442
450
480
490
510
520
Parameter
Acid cap. 01758
Aluminium 14825
Aluminium 00594
BOD 00687
Bromine 00605
Calcium 00858
Chlorine dioxide
Colour
HydroPerox 18789
Iodine 00606
Magnesium 00815
Molybdenum 00860
Monochloramine
Nitrate 14556
Oxygen 14694
Oxyg. scavengers
Ozone 00607
pH 01744
Potassium 14562
Potassium 00615
Residual hardness 14683
Sulfate 14548
Sulfate 00617
Sulfate 14564
Sulfide 14779
Suspended solids
Tin 14622
Total hardness 00961
Turbidity
Chapter 5.7 | Page 1 of 4

Advertisement

Table of Contents
loading

Table of Contents