The display now shows:
Enter the concentration of the second standard;
e.g.: [0] [,] [1] for 0.1.
• Back with the key [Esc].
• EBackout the entry with the key [F1].
Confirm your selection by pressing [Enter].
The display now shows:
Prepare the second standard and press [Test].
The display shows the entered value and the measured
absorbance value:
Confirm your selection by pressing [Enter].
Note
• To measure further standards, follow the above procedure.
• At least two standards must be measured.
• A maximum of 14 standards (S1 to S14) can be measured.
When the desired number of standards or the maximum num-
ber of 14 standards have been measured, press the key [Store].
S1 accepted
S2: +_____
ESC
F1
S2: 0.10 mg/l
Prepare
Press TEST
S2: 0.10 mg/l
mAbs: 150
Chapter 5.6 | Page 5 of 13