1.2.1 Lithium powerpack –
Important information
The lithium powerpack (
(saved measurement results, settings) for approx. ½ year when
the unit is not being supplied with power by the battery pack
or the mains supply.
As long as the colorimeter is being supplied with mains or
battery power, it does not draw any power from the lithium
powerpack. Since lithium powerpacks have a very long life,
it will in all probability not be necessary to exchange the
one supplied. All the same, to ensure proper functioning it is
recom mended to replace the lithium powerpack every five years.
Note
When neither the mains adapter nor the battery pack are
supplying power, removal of the lithium powerpack will result
in a complete loss of data (saved measurement results and
settings). For this reason we strongly recommend that the unit
is connected to the mains power supply while replacing the
lithium powerpack.

1.2.2 Charging the battery pack

The rechargeable batteries (
m eter for recharging. Once the mains power supply is established,
the batteries start to be charged.
Charge the rechargeable batteries in the instrument for 5 days
(working with power from mains is possible). Now use the
instrument without mains until the first battery warning comes
up. Charge again, 4 days this time. Repeat usage to battery
warning and charging four times.
Approximately 10 charging/discharging cycles are required until
the batteries reach their full capacity.
The unit can be operated with mains power supply with or
without the batteries in place. For this you must attach the
country-specific wall-socket adapter to the mains adapter.
) serves to save all important data
B
) remain in place in the photo-
A
en
9