Note
The running countdown can be skipped by pressing the [Enter]
key once. In this case the user must observe the necessary
reaction time him-/herself. (Failure to observe the specified
reaction time can lead to erroneous results.) Further options
to deactivate the countdown procedure: mode No. 13 or
Profi mode (mode No. 50).
After the countdown has expired the display shows:
Place the prepared blank in the measurement compartment
with the mark on the cell pointing towards the mark on
the unit case.
Positioning the cell (ø 24 mm)
Align the triangular mark on the cell with that on the Spectroquant®
Multy:
To afford better protection against sunlight, press the o-ring firmly
into place.
90 Bromine 00605
0.10-5.00 mg/l Br2
prepare Zero
press ZERO