Activating/Deactivating The Countdown Function (Observance Of Reaction Times) - Millipore Spectroquant Multy Operating Manual

Colorimeter
Table of Contents

Advertisement

Pressing the [0] key deactivates the signal beep function.
Press the [1] key activates the signal beep function.
Confirm by pressing [Enter].
Note
In connection with methods that include a reaction time, the
unit emits an acoustic signal in the last 10 seconds before the
countdown expires even when the signal beep function is inactive.
3.3.3 Activating/deactivating the countdown
function (observance of reaction times)
Certain methods require the observance of reaction times.
These waiting times are stored as standard settings in the
respective methods in the form of a timer (countdown)
function. The countdown can be deactivated for all methods
involved in the following manner:
Press the keys [Mode] [1] [3] in succession.
Confirm your selection by pressing [Enter].
The display shows:
Pressing the [0] key deactivates the countdown function.
Pressing the [1] key activates the countdown function.
Confirm by pressing [Enter].
Note
When the countdown function is inactive, the necessary
reaction time must be observed by the user him-/herself. Failure
to observe the specified reaction time can lead to erroneous
results.
<countdown>
actual:
switched on
ON:1
OFF:0
en
37